Table 4 Adverse events encountered during the procedure

From: Treatment of malignant airway obstruction with Y-shape sigma stent loaded with I125 seeds installed via rigid bronchoscopy

Adverse events

# of case

Potential cause(s)

Incomplete expansion of the stent

3

Stent was stiff, tumor suppression

Unsatisfied location of the stent

4

The two limbs of Y-shape stent were too short or in same length, difficult to insert or pass through the stenotic location

Stent migration

2

Enlarged airway lumen after ablative surgery, size of the stent was small

125I seeds loss

2

The coating was hard and brittle, and torn out due to repeated expanding and collapsing

Guide wire broken

1

Incorrect operation

Rupture of coating

3

Incorrect operation

Rupture of bronchi

2

Whole stent was placed into only one side of the main bronchus

Bleeding

12

Tumor angiogenesis and its blood supply

vessels were not blocked by embolizing pulmonary artery before the procedure

Pneumothorax

1

Rupture of a bronchus lead to mediastinal pneumothorax